戴安公司生產(chǎn)的儀器主要有離子色譜儀、高效液相色譜儀、快速溶劑萃取儀和氨基酸直接分析儀。其中離子色譜儀憑借著雄厚的技術(shù)優(yōu)勢和方法已廣泛地應(yīng)用于各個領(lǐng)域,其*達(dá)80%。與離子色譜相關(guān)的文章有90%均出自戴安公司生產(chǎn)的離子色譜儀。
型號 | 美國戴安 P/N: 075402 |
---|---|
廠商性質(zhì) | 代理商 |
更新日期 | 2018-04-16 |
The IonPac? CS16 is the column of choice for disparate concentration ratios of adjacent eluting cations such as sodium and ammonium in diverse sample matrices. The high-capacity, high-resolution CS16 column can be used for short-chained amines, including alkylamines and alkanolamines, in diverse sample matrices. The CS16 column is now available in capillary format (0.5 × 250 mm) offering reduced eluent consumption and reduced operating costs.
Using an isocratic acid eluent and elevated temperature (40 °C) coupled with suppressed conductivity, ratios up to 10,000:1 of sodium and ammonium can be resolved in less than 20 min. Sample matrices include environmental waters; power plant waters treated with ammonium, morpholine or ethanolamine; chemical additives; chemical process solutions; scrubber solutions; plating baths; and industrial solvents. The CS16 is designed for use in most IonPac CS15 applications.
The CS16 column has the highest capacity among Dionex carboxylate-based cation columns, resulting in improved loadability and resolution. Use the CS16 with the Eluent Generator for simplified, consistent methanesulfonic acid eluent preparation. Use the Cation Self-Regenerating Suppressor (CSRS? 300) with standard or microbore CS16 columns or the Cation Capillary Electrolytic Suppressor (CCES? 300) with the CS16 capillary column.
IonPac CS16 Cation-Exchange Column Specifications | |
---|---|
Dimensions | IonPac? CS16 Capillary Column: 0.5 × 250 mm IonPac CG16 Capillary Guard Column: 0.5 × 50 mm IonPac CS16 Analytical Column: 5 × 250 mm and 3 × 250 mm IonPac CG16 Guard Column: 5 × 50 mm and 3 × 50 mm |
Maximum Operating Pressure | 4000 psi (Standard and Microbore), 5000 psi (Capillary) |
Phase Compatibility | Acidic eluents; 0–100% HPLC acetonitrile. Alcohols should be avoided. |
Substrate Characteristics | Bead Diameter: 5 ?m Cross-Linking (%DVB): 55% |
Functional Group Characteristics | Ion-Exchange Groups: Grafted carboxylic acid Surface Characteristics: Medium hydrophobic |
Capacity | 84 ?eq (0.5 × 250 mm capillary column) 17 ?eq (0.5 × 50 mm capillary guard column) 8400 ?eq (5 × 250 mm analytical column) 1700 ?eq (5 × 50 mm guard column) 3000 ?eq (3 × 250 mm analytical column) 600 ?eq (3 × 50 mm guard column) |
Column Construction | PEEK? with 10-32 threaded ferrule-style end fittings. All components are nonmetallic. |
Analytical Columns | |
---|---|
IonPac CS16 Capillary Column (0.5 x 250 mm) | 075401 |
IonPac CS16 Analytical Column (5 x 250 mm) | 079805 |
IonPac CS16 Analytical Column (3 x 250 mm) | 059596 |
Guard Columns | |
---|---|
IonPac CG16 Capillary Guard Column (0.5 x 50 mm) | 075402 |
IonPac CG16 Guard Column (5 x 50 mm) | 057574 |
IonPac CG16 Guard Column (3 x 50 mm) | 079931 |